11.1 Routine Cleaning and Disinfection
Cleaning the mattress:
►
Check mattress cover top for any signs of damage.
Replace or repair and completely disinfect mattress cover top if damaged.
►
Check inside of mattress cover top for signs of liquid ingress.
Replace or clean and completely disinfect mattress cover top if damp inside.
►
Leave mattress cover on mattress.
►
Clean with 60°C warm water and cleaning detergent.
►
Rinse mattress with cold water.
►
Let mattress dry or wipe dry.
►
Wipe mattress with disinfectant.
►
Rinse mattress with cold water.
►
Let mattress dry or wipe dry.
Cleaning the SCU:
►
Wipe SCU with disinfectant.
►
Let SCU dry or wipe dry.
11.2 Full Cleaning and Disinfection
Cleaning the mattress:
►
Deflate mattress and remove cover (see Removing the Mattress Cover).
►
Check mattress cover top and base for any signs of damage.
Replace or repair and completely disinfect mattress cover top and base if damaged.
►
Check mattress cover top and base for signs of liquid ingress.
Replace or clean and completely disinfect mattress cover top and base if damp inside.
►
Clean all mattress cells and pipes with 60°C warm water and cleaning detergent.
►
Rinse mattress with cold water.
►
Let mattress dry air dry or wipe dry.
►
Wipe mattress with disinfectant.
►
Rinse mattress with cold water.
►
Let mattress dry air dry or wipe dry.
Cleaning the mattress cover:
►
Remove cover (see Removing the Mattress Cover).
►
If machine washing mattress top/base covers, the temperature should be raised during the wash cycle, to 65°C/149°F, for
10 -15 minutes, or 71°C/160° F, for 3 - 10 minutes, using hospital approved detergents and rinsing agents.
NOTE
Maximum wash temperature is 75°C/167°F.
►
Dry cover in tumble dryer at low temperature.
Cleaning the air pipe:
►
Wipe air pipe with cleaning agent or disinfectant.
►
Let air pipe dry.
Cleaning the SCU:
►
Remove filter.
►
Wipe SCU and filter with disinfectant.
►
Let SCU and filter dry.
►
Reinsert filter.