![LINET OptiCare X Instructions For Use And Technical Description Download Page 22](http://html1.mh-extra.com/html/linet/opticare-x/opticare-x_instructions-for-use-and-technical-description_1922864022.webp)
D9U003VSX-0101_00
22
5.4.2 Manufacturer Instructions - Electromagnetic Susceptibility
Immunity Tests
Compliance level
Electrostatic discharge (ESD)
IEC 61000-4-2
± 8 kV for contact discharge
± 15 kV for air discharge
Radiated RF
IEC 61000-4-3
Proximity fields from RF wireless communications equipment
IEC 61000-4-3
3 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
See Table 1
Fast electrical transients / burst
IEC 61000-4-4
±2 kV for power line
repetition frequency 100 kHz
Surge
IEC 61000-4-5
± 1 kV Line-to-line
± 2 kV Line-to-ground
Conducted RF
IEC 61000-4-6
3 V (0,15 MHz – 80 MHz)
6 V in ISM bands between 0,15 MHz and 80 MHz
80 % AM at 1 kHz
Power frequency (50/60 Hz) magnetic field
IEC 61000-4-8
30 A/m
Voltage dips, short interruptions on power supply input lines
IEC 61000-4-11
0 % UT; 0,5 cycle
At 0°, 45°, 90°, 135°, 180°, 225°, 270° a 315°
0 % UT; 1 cycle and 70 % UT; 25/30 cycle
Single phase: at 0°
0 % UT; 250/300 cycle
Table 1 - IMMUNITY to RF wireless communications equipment
Test frequency (MHz)
Band (MHz)
Service
Modulation
Immunity Test Level
V/m
385
380 - 390
TETRA 400
Pulse modulation 18 Hz
27
450
430 - 470
GMRS 460, FRS 460
FM ± 5 kHz deviation 1
kHz sine
28
710
745
780
704 - 787
LTE band 13, 17
Pulse modulation 217
Hz
9
810
870
930
800 - 960
GSM 800/900, TETRA
800, iDEN 820, CDMA
850,
LTE band 5
Pulse modulation 18 Hz
28
1 720
1 845
1 970
1 700 - 1 990
GSM 1800; CDMA
1900; GSM 1900;
DECT; LTE band 1, 3,
4, 25; UMTS
Pulse modulation 217
Hz
28
2 450
2 400 - 2 570
Bluetooth, WLAN,
802.11 b/g/n, RFID
2450,
LTE band 7
Pulse modulation 217
Hz
28
5 240
5 500
5 785
5 100 - 5 800
WLAN 802.11 a/n
Pulse modulation 217
Hz
9
NOTE
There are applied no deviations to requirements of IEC 60601-1-2 ed. 4
NOTE