D9U003VSX-0101_00
20
5
Technical Specification (OptiCare X)
All technical data are rated data and are subject to construction and manufacturing tolerances.
WARNING!
If Multicare X bed is used with OptiCare X integrated mattress system, respect values of mechanical and elect
-
rical specifications which can harm none of them!
5.1
Mechanical Specifications
Parameter
Value
External dimensions of Inflated Mattress (length x width x height)
214 cm
x
87 cm
x
23 cm
Length of the Integrated Mattress Extension
19 cm
Dimensions of the X-ray Pocket (length x width)
90 cm
x
92 cm
External dimensions of SCU
36 cm
x
22 cm
x
10 cm
Weight of Inflated Mattress
14 kg (31 lbs)
Weight of SCU
6 kg (13.1 lbs)
Inflation time after storage
max. 10 min
CPR deflation time (depending on patient weight, chosen mode - optimization
or maximum internal pressure - and on type of CPR - electric or manual)
max. 30 s
Remain Inflated in Transport Mode
12 hours (whem starting from Maximum
Inflate Mode)
Sound Pressure Level
max. 46 dBA (normal operation without
any sounding alert signal)
Maximum Patient Weight
250 kg (551 lbs)
Minimum Mattress Load
40 kg
Maximum Washing Temperature of the mattress cover ELITE (MIC200)
71°C
Maximum Washing Temperature of the mattress cover ENDURANCE (SIOEN
RP9774-2)
71°C
Maximum Washing Temperature of the mattress cover ZONED (DARTEX)
55°C
Maximum Drying Temperature of the mattress cover ELITE (MIC200)
55°C
Maximum Drying Temperature of the mattress cover ENDURANCE (SIOEN
RP9774-2)
55°C
Maximum Drying Temperature of the mattress cover ZONED (DARTEX)
55°C
Fire Resistance of the Cover
according to BS7175 - ignition source 5
Fire Resistance of the Top Cover
according to 16 CFR 1632
Fire Resistance of the whole mattress
according to BS7177 - ignition source 5,
16 CFR 1632
5.2
Electrical Specifications
Parameter
Value
Input Voltage, Frequency:
EU version
USA version
Brazilian version
220-240 V AC, 50/60 Hz
100-127 V AC, 50/60 Hz
220 V AC, 60 Hz
Maximum Power Input
max. 40 VA (when operating from mains supply)
Ingress Protection (EN 60529)
IPX4
Protection Class
Class I with Applied Parts Type B
Electrical Safety
in conformity with EN 60601-1
SCU Fuse
220-240 V AC version
100-127 V AC version
2 x T0,5AH Anti-surge Fuse
2 x T1AH Anti-surge Fuse