![LINET OptiCare X Instructions For Use And Technical Description Download Page 34](http://html1.mh-extra.com/html/linet/opticare-x/opticare-x_instructions-for-use-and-technical-description_1922864034.webp)
D9U003VSX-0101_00
34
OptiCare X for Multicare X
MOBILE
PRONE
MAX
OPTIMIZE
25. 12. 25
22:58
OFF
LOW
RE-OPTIMIZE
HIGH
MICROCLIMATE
78
,5 kg
9.4.4 OPTIMIZE MODE
When OptiCare X has finished inflating, it will switch to Optimize Mode. This is the default mode. As long as the patient remains
on the mattress automatic optimization of the mattress internal pressure continues. Optimization will occur if the patient´s position
changes sufficiently to trigger Optimization Detection, or if initiated by the Optimization automatic timer. The integrated MICROC
-
LIMATE Management system will start working automatically when the patient gets into bed, and stops if the patient gets out
(MICROCLIMATE OFF).
OptiCare X system in OPTIMIZE Mode detects when a patient has entered or left the bed. This automatically starts the optimization
process on patient ingress and puts the mattress into Standby mode on patient egress. During Standby mode mattress areas A and
B are inflated to a static pressure. There is a short stable pressure detection delay before reacting to change of status to prevent
unnecessary mode changes because patient has changed position. If at any time nursing staff feel it necessary to re-optimize the
patient then this can be initiated manually by touching the RE-OPTIMIZE Icon.
Optimization will stop working, the mattress will be inflated and air pressure level will be set according to the patient‘s
weight if:
►
The mattress support platform is tilted by 10 and more degrees (Anti-Trendelenburg Tilt).
►
The mattress support platform is tilted by 10 and more degrees (Lateral Tilt).
►
The mattress support platform is tilted by 5 and more degrees (Trendelenburg Tilt).
►
Tilt reduction of the mattress support platform has value of 7 or less degrees (Anti-Trendelenburg Tilt).
►
Tilt reduction of the mattress support platform has value of 7 or less degrees (Lateral Tilt).
►
Tilt reduction of the mattress support platform has value of 3 or less degrees (Trendelenburg Tilt).
►
Automatic Lateral Therapy (ALT) is activated.
To perform manual re-optimization:
►
press RE-OPTIMIZE Icon.
To set intensity of MICROCLIMATE Management:
►
press MICROCLIMATE OFF Icon to turn off the MICROCLIMATE Management.
►
press MICROCLIMATE LOW Icon to lower intensity of the MICROCLIMATE Management.
►
press MICROCLIMATE HIGH Icon to intensify the MICROCLIMATE Management.
NOTE
: MICROCLIMATE Management will be automatically OFF if there is no patient on the mattress.
Fig. OPTIMIZE MODE