Operating Counter
248083389-1 Rev
6-5
•
LOCATION SELECT button:
Allows user to change location
before sampling.
•
RECIPE button:
Allows the user to view, load and unload recipes
that have been configured and saved in the recipe data base.
Note:
Hard copy
printouts are available only
when the optional thermal
printer is installed.
•
PRINT LAST RECORD/SAVE LAST RECORD TO FLASH
DRIVE:
Prints the last recorded sample using the current
configuration to determine the type of data printed (i.e. cumulative
vs. differential, raw vs. normalized, ft
3
vs. m
3
). When the USB
option is enabled, the last record is saved to a flash drive.
•
The print and USB options are configured through the OUTPUT
button in the CONFIG screen. For more details about printing, see
the Configuration section later in this chapter.
•
Date/Time:
Displays the current date and time.
Note:
If the instrument is
counting when the “X”
appears in the battery
indicator, the pump will
automatically stop to prevent
the battery from discharging
completely.
If this happens, attach the unit
to AC and make sure the front
power switch is ON.
The unit will recharge
completely in about 4-5 hours.
Users can continue to run
samples and take counts
while the unit is plugged in.
•
Battery Indicator:
Indicates that the instrument is being powered
by a rechargeable battery. The amount of battery life left is denoted
by the fill inside the battery icon. When the battery is low, the
words “BATT LOW!” will appear on the screen and the unit will
beep continually until it is plugged into its AC power cord for
recharging. See Figure 6-4.
Figure 6-4 Battery Level Indicators
Full
Depleted
Summary of Contents for solair 3100
Page 1: ......
Page 2: ...00...
Page 3: ...Lighthouse Worldwide Solutions S3100 S5100 Boulder Counter Gen E Operating Manual...
Page 6: ......
Page 12: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual t vi 248083389 1 Rev...
Page 14: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual ii 248083389 1 Rev...
Page 19: ...Introduction 248083389 1 Rev 2 3 Figure 2 1 SOLAIR Dimensions Side Dimensions Top Dimensions...
Page 22: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual 2 6 248083389 1 Rev...
Page 36: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual 3 14 248083389 1 Rev...
Page 60: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual 4 24 248083389 1 Rev...
Page 66: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual 5 6 248083389 1 Rev...
Page 115: ...Operating Counter 248083389 1 Rev 6 49 Figure 6 64 SOLAIR 3100 STATUS Screen...
Page 180: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual D 2 248083389 1 Rev...
Page 187: ...00...
Page 188: ......