MODBUS Register v1.48
248083389-1 Rev
B-17
Threshold Setup Registers
Threshold data is stored in the input registers in the 45xxx series which
are read/write. All threshold data items are 4 bytes long and are stored
across 2 registers. Byte and word ordering is big-endian. Thus, data
items are formed by placing the high bytes in front of the low bytes.
For example:
<High Bytes><Low Bytes> = <4 Byte Data Item>
For particle channels, the threshold value is a 32-bit unsigned integer.
If the data value exceeds the threshold value and the alarm is enabled
for that channel, the threshold flag in the Data Status register (30007-
30008, bit 4) is set.
Note:
The table below
shows the registers for an
8 channel particle counter.
Counters with fewer
channels do not use the
extra registers. The
smallest particle channel
starts at the xxx09 position.
The threshold registers (45xxx series) run in parallel with the data
registers (30xxx series). For example, data register 30010’s
corresponding threshold register would be 45010. Data register 30016’s
threshold register would be 45016.
Table B-16 Alarm Threshold Registers
Register Data
Type
Description
45009
unsigned int Threshold for Particle Channel 1 [high]
(smallest particle size starts here)
45010
unsigned int Threshold for Particle Channel 1 [low]
45011
unsigned int Threshold for Particle Channel 2 [high]
45012
unsigned int Threshold for Particle Channel 2 [low]
45013
unsigned int Threshold for Particle Channel 3 [high]
45014
unsigned int Threshold for Particle Channel 3 [low]
45015
unsigned int Threshold for Particle Channel 4 [high]
45016
unsigned int Threshold for Particle Channel 4 [low]
45017
unsigned int Threshold for Particle Channel 5 [high]
45018
unsigned int Threshold for Particle Channel 5 [low]
45019
unsigned int Threshold for Particle Channel 6 [high]
45020
unsigned int Threshold for Particle Channel 6 [low]
45021
unsigned int Threshold for Particle Channel 7 [high]
45022
unsigned int Threshold for Particle Channel 7 [low]
Summary of Contents for solair 3100
Page 1: ......
Page 2: ...00...
Page 3: ...Lighthouse Worldwide Solutions S3100 S5100 Boulder Counter Gen E Operating Manual...
Page 6: ......
Page 12: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual t vi 248083389 1 Rev...
Page 14: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual ii 248083389 1 Rev...
Page 19: ...Introduction 248083389 1 Rev 2 3 Figure 2 1 SOLAIR Dimensions Side Dimensions Top Dimensions...
Page 22: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual 2 6 248083389 1 Rev...
Page 36: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual 3 14 248083389 1 Rev...
Page 60: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual 4 24 248083389 1 Rev...
Page 66: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual 5 6 248083389 1 Rev...
Page 115: ...Operating Counter 248083389 1 Rev 6 49 Figure 6 64 SOLAIR 3100 STATUS Screen...
Page 180: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual D 2 248083389 1 Rev...
Page 187: ...00...
Page 188: ......