Operating Counter
248083389-1 Rev
6-73
Setting up Counter to Run Reports
For this example, the FED STD FT
3
report will be used. See
Figure 6-95.
Figure 6-95 FED STD FT3 Report Screen
•
Set the instrument for the Fed Std ft
3
report, 100 ft
2
, Class 100,
unidirectional air flow, operational room status.
•
This will generate the following minimum requirements:
Note:
The Minimum
Volumes are measured in the
standard measurement
required by each specific
report.
The number of digits allowed
for area to calculate volume is
limited to a 5 digit value for ft2
and a 4 digit value for m2.
•
Min Locations: 4
•
Min Samples: 5
Table 6-1 Table - Minimum Volume per Channel Size
Ch Size (
μ
)
Min Vol (ft
3
)
0.3
0.100
0.5
0.200
1.0
0.917
3.0
10.309
5.0
31.696
10.0
145.985
Summary of Contents for solair 3100
Page 1: ......
Page 2: ...00...
Page 3: ...Lighthouse Worldwide Solutions S3100 S5100 Boulder Counter Gen E Operating Manual...
Page 6: ......
Page 12: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual t vi 248083389 1 Rev...
Page 14: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual ii 248083389 1 Rev...
Page 19: ...Introduction 248083389 1 Rev 2 3 Figure 2 1 SOLAIR Dimensions Side Dimensions Top Dimensions...
Page 22: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual 2 6 248083389 1 Rev...
Page 36: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual 3 14 248083389 1 Rev...
Page 60: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual 4 24 248083389 1 Rev...
Page 66: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual 5 6 248083389 1 Rev...
Page 115: ...Operating Counter 248083389 1 Rev 6 49 Figure 6 64 SOLAIR 3100 STATUS Screen...
Page 180: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual D 2 248083389 1 Rev...
Page 187: ...00...
Page 188: ......