Lighthouse S3100/S5100/Boulder Counter Operating Manual
7-8
248083389-1 Rev
MANUAL Counting Mode
In Manual counting mode, the sample time is based on when the
counter is instructed to stop counting. At this point, a data record is
recorded and the sample time is the interval between the command to
start counting and the command to stop counting.
7
Start Pump
9
Start Manual Count
10
Stop Manual Count; writes a data record, uses the time interval as
the sample time
The "hold" time is effectively the time between a STOP and a START
command.
After the last desired sample is taken, send the following to stop the
pump:
8
Stop Pump
AUTOMATIC Counting Mode
In Automatic counting mode, the instrument uses the configured
sample time, hold time, and initial sample delay to record samples.
The instrument will continue running samples until it receives a stop
command. When the stop command is given, since the device will not
have completed a full sample cycle, the most current data will not
record to the buffer.
After setting all the instrument parameters, run these commands:
11
Start Instrument; to start recording
12
Stop Instrument; to stop recording
Summary of Contents for solair 3100
Page 1: ......
Page 2: ...00...
Page 3: ...Lighthouse Worldwide Solutions S3100 S5100 Boulder Counter Gen E Operating Manual...
Page 6: ......
Page 12: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual t vi 248083389 1 Rev...
Page 14: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual ii 248083389 1 Rev...
Page 19: ...Introduction 248083389 1 Rev 2 3 Figure 2 1 SOLAIR Dimensions Side Dimensions Top Dimensions...
Page 22: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual 2 6 248083389 1 Rev...
Page 36: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual 3 14 248083389 1 Rev...
Page 60: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual 4 24 248083389 1 Rev...
Page 66: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual 5 6 248083389 1 Rev...
Page 115: ...Operating Counter 248083389 1 Rev 6 49 Figure 6 64 SOLAIR 3100 STATUS Screen...
Page 180: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual D 2 248083389 1 Rev...
Page 187: ...00...
Page 188: ......