Lighthouse S3100/S5100/Boulder Counter Operating Manual
6-24
248083389-1 Rev
SETTING
The Settings screen allows the user to configure the instrument to count
in different modes and formats when running a sample. See
Figure 6-30.
Figure 6-30 Sample Settings Screen
COUNT MODE
The following modes are available: Auto, Manual, Beep and
Concentration.
•
AUTO
- When the instrument is in Automatic Mode and the
START button is pressed, the instrument will start counting
particles automatically according to the Sample Time, Hold Time
and the number of Cycles that are configured.
If Cycles are set to 0, the instrument will continue indefinitely in
Auto Mode until the STOP button is pressed.
•
MANU (Manual Mode)
- When the instrument is in Manual
Mode, it will start counting when START is pressed and stop at the
end of one programmed Sample Time.
•
CONC (Concentration Mode)
- When the instrument is in
Concentration mode, it gives a calculated value of the concentration
of particles in a volume of air, measured and displayed on the
MAIN screen in either counts per cubic foot or per cubic meter.
Note:
The sample time
for Auto, Manual and Beep
modes count down and the
sample times for
Concentration mode count
up.
Counting starts when the START button is pressed and it will
continue until the STOP button is pressed. The sample time for
Concentration mode is six seconds. As the sample time on the
MAIN screen counts from one to six, the particle counts are
updated continuously.
Summary of Contents for solair 3100
Page 1: ......
Page 2: ...00...
Page 3: ...Lighthouse Worldwide Solutions S3100 S5100 Boulder Counter Gen E Operating Manual...
Page 6: ......
Page 12: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual t vi 248083389 1 Rev...
Page 14: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual ii 248083389 1 Rev...
Page 19: ...Introduction 248083389 1 Rev 2 3 Figure 2 1 SOLAIR Dimensions Side Dimensions Top Dimensions...
Page 22: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual 2 6 248083389 1 Rev...
Page 36: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual 3 14 248083389 1 Rev...
Page 60: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual 4 24 248083389 1 Rev...
Page 66: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual 5 6 248083389 1 Rev...
Page 115: ...Operating Counter 248083389 1 Rev 6 49 Figure 6 64 SOLAIR 3100 STATUS Screen...
Page 180: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual D 2 248083389 1 Rev...
Page 187: ...00...
Page 188: ......