Setting up the Counter
248083389-1 Rev
4-3
The set up process is as follows:
Analog channels are labeled in the settings as A1 through A4. A1 and
A2 correspond to Analog port
1
on the chassis and A3, A4 correspond
to port
2
. A1 and A3 are the first channel and A2 and A4 are the second
channel of these ports. So, if a single-channel sensor is attached to port
1
and is wired to connect to channel one, its data will be displayed on
A1. If it is wired to operate on the second channel, its data will be
displayed on A2.
If a dual-channel sensor, such as the Lighthouse TRH (temperature,
relative humidity) probe, is connected to port
1
, its temperature data
will show up on A1 and its relative humidity data will be displayed on
A2.
The setup process is as follows:
Note:
The screen
examples in this chapter
are for visual reference.
The values displayed may
not be indicative of actual
values.
1. From the MAIN Screen, press the CONFIG button.
2. On the CONFIGURATION screen, press the ANALOG button.
Note:
The Min and Max
values for the Lighthouse
TRH Probe are 0-150. If
using a TRH Probe other
than Lighthouse, please
check the manufacturer’s
specifications for the
correct values.
3. The Analog Configuration Screen displays Figure 4-2.
Figure 4-2 Analog Configuration Screen
4. Press the red “
X
” button to enable an analog channel.
Summary of Contents for solair 3100
Page 1: ......
Page 2: ...00...
Page 3: ...Lighthouse Worldwide Solutions S3100 S5100 Boulder Counter Gen E Operating Manual...
Page 6: ......
Page 12: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual t vi 248083389 1 Rev...
Page 14: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual ii 248083389 1 Rev...
Page 19: ...Introduction 248083389 1 Rev 2 3 Figure 2 1 SOLAIR Dimensions Side Dimensions Top Dimensions...
Page 22: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual 2 6 248083389 1 Rev...
Page 36: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual 3 14 248083389 1 Rev...
Page 60: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual 4 24 248083389 1 Rev...
Page 66: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual 5 6 248083389 1 Rev...
Page 115: ...Operating Counter 248083389 1 Rev 6 49 Figure 6 64 SOLAIR 3100 STATUS Screen...
Page 180: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual D 2 248083389 1 Rev...
Page 187: ...00...
Page 188: ......