Operating Counter
248083389-1 Rev
6-7
•
SAMPLE:
The Sample Time (hh:mm:ss) is the duration of one
counting cycle. The Sample Time will count down on the MAIN
screen when the instrument is in AUTO or Manual mode so the user
can see how much time is remaining in the sample period. In
Concentration mode, the Sample Time will count up to 6 seconds
per cycle.
Note:
If Hold time is
greater than 1 minute, the
pump will stop during that
time. At the end of the hold
time, the pump will restart.
•
HOLD:
Displays the hold time in between cycles. The maximum
hold time is 99 hours, 59 minutes, 59 seconds.
•
RECS:
This displays the current number of records stored in the
instrument and the total number of records that can be stored. The
data buffer is a circular buffer. The instrument can store up to 3000
records. An asterisk (*) will appear in front of counts when the
buffer wraps.
•
START/STOP:
Press START button on the screen to start
counting. When running, the instrument will display
“COUNTING” in the lower right corner of the screen as shown in
Figure 6-5. Press the STOP button to stop counting; the word
“STOPPED” will display.
Figure 6-5 Counting Mode
Summary of Contents for solair 3100
Page 1: ......
Page 2: ...00...
Page 3: ...Lighthouse Worldwide Solutions S3100 S5100 Boulder Counter Gen E Operating Manual...
Page 6: ......
Page 12: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual t vi 248083389 1 Rev...
Page 14: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual ii 248083389 1 Rev...
Page 19: ...Introduction 248083389 1 Rev 2 3 Figure 2 1 SOLAIR Dimensions Side Dimensions Top Dimensions...
Page 22: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual 2 6 248083389 1 Rev...
Page 36: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual 3 14 248083389 1 Rev...
Page 60: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual 4 24 248083389 1 Rev...
Page 66: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual 5 6 248083389 1 Rev...
Page 115: ...Operating Counter 248083389 1 Rev 6 49 Figure 6 64 SOLAIR 3100 STATUS Screen...
Page 180: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual D 2 248083389 1 Rev...
Page 187: ...00...
Page 188: ......