Lighthouse S3100/S5100/Boulder Counter Operating Manual
6-64
248083389-1 Rev
REPORTS
The SOLAIR can print the following Cleanroom Classification to
Standards reports:
Note:
The EU GMP 2009
report is only valid if the
instrument is configured with
the 0.5 and 5.0 micron
channel sizes.
Federal Standard ft
3
ISO 14644-1
EU GMP 2009
Reports General Settings
Note:
If the user sets the
counter for a minimum
volume, and takes the exact
number of minimum locations
and samples, the user could
get a “Cannot Classify: too
few locations” error if the
minimum volume rounds the
sample time down so that not
enough samples at the
minimum volume are taken.
In this case, add one or two
seconds to the sample time
so that the user is not at the
exact minimum volume.
Enable the Reports feature by pressing the red “
X
” (Default position)
located at the top left of the screen. The button will display a check
mark indicating the feature is enabled. See Figure 6-80.
Figure 6-80 Typical Reports Screen
•
Depending on the reports requirements, the number of cycles and
minimum sample time will be calculated when the reports mode is
enabled.
•
The number of required locations will be automatically assigned to
the report. The user may change locations via the reports SETUP
screen.
•
The display format will be changed to Normalized (normalized to
ft3 for FED-209 and m3 for ISO and EU GMP). The user may
change the display mode from the reports SETTINGS screen.
Summary of Contents for solair 3100
Page 1: ......
Page 2: ...00...
Page 3: ...Lighthouse Worldwide Solutions S3100 S5100 Boulder Counter Gen E Operating Manual...
Page 6: ......
Page 12: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual t vi 248083389 1 Rev...
Page 14: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual ii 248083389 1 Rev...
Page 19: ...Introduction 248083389 1 Rev 2 3 Figure 2 1 SOLAIR Dimensions Side Dimensions Top Dimensions...
Page 22: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual 2 6 248083389 1 Rev...
Page 36: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual 3 14 248083389 1 Rev...
Page 60: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual 4 24 248083389 1 Rev...
Page 66: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual 5 6 248083389 1 Rev...
Page 115: ...Operating Counter 248083389 1 Rev 6 49 Figure 6 64 SOLAIR 3100 STATUS Screen...
Page 180: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual D 2 248083389 1 Rev...
Page 187: ...00...
Page 188: ......