Lighthouse REMOTE 3104P, 5104P Operating Manual
4-2
248083336-1 Rev 3
DIP Switch
Settings
GENERAL DEFINITIONS
OFF (LEFT) = 0, ON (RIGHT) = 1
Note:
The DIP Switches
must be set before the unit
is powered on.
Communications Mode (DIP #8)
In order to communicate with the instrument using the MODBUS
protocol, DIP switch 8 must be set to OFF.
REMOTE Display Mode (DIP#7)
In order to use the REMOTE Display option, DIP switch 7 must be set
to ON and the Display must be connected to the instrument before it is
turned on.
WARNING:
Always
turn the instrument OFF
before connecting or
disconnecting the
REMOTE Display.
If DIP switch 7 is ON but the REMOTE Display is not connected when
the instrument is powered on, the unit will not start. If the Display is
not needed, power the instrument OFF, change DIP switch 7 to OFF
and power the instrument ON. Otherwise, power the instrument OFF,
connect the REMOTE Display then power the instrument ON.
Table 4-1 DIP Switch settings
Position#
Description
Setting
1
Binary Bit 0
OFF=0, ON=1
2
Binary Bit 1
OFF=0, ON=1
3
Binary Bit 2
OFF=0, ON=1
4
Binary Bit 3
OFF=0, ON=1
5
Binary Bit 4
OFF=0, ON=1
6
Binary Bit 5
OFF=0, ON=1
7
Remote Display
OFF=0, ON=1
8
Communications
Mode
OFF: MODBUS
Summary of Contents for REMOTE 3104P
Page 1: ......
Page 2: ...00...
Page 3: ...Lighthouse Worldwide Solutions REMOTE 3104P 5104P Airborne Particle Counter Operating Manual...
Page 6: ......
Page 10: ...Lighthouse REMOTE 3014P 5104P Operating Manual t iv 248083336 1 Rev 3...
Page 12: ...Lighthouse REMOTE 3014P 5104P Operating Manual ii 248083336 1 Rev 3...
Page 40: ...Lighthouse REMOTE 3104P 5104P Operating Manual 4 10 248083336 1 Rev 3...
Page 54: ...Lighthouse REMOTE 3104P 5104P Operating Manual 6 12 248083336 1 Rev 3...
Page 74: ...Lighthouse REMOTE 3104P 5104P Operating Manual B 2 248083336 1 Rev 3...
Page 76: ...Lighthouse REMOTE 3104P 5104P Operating Manual C 2 248083336 1 Rev 3...
Page 81: ...00...
Page 82: ......