Lighthouse REMOTE 3104P, 5104P Operating Manual
5-2
248083336-1 Rev 3
3.
Configure the unit via the user interface to sample for 30 minutes.
4.
Allow the instrument to sample through a 30 minute period. This
time allows the unit to warm up and purge any residual particles
that might be inside it.
5.
Configure the unit via the user interface to sample for 5 minutes
and set a 10 second hold.
6.
Set Cycles to 10 so the instrument will take 10 five-minute
samples.
7.
If an average of more than one count per five minute sample is
reported, reset the instrument to sample for 30 minutes again to
purge it, then repeat the Purge Test again.
8.
After the instrument has met the requirement of the Purge test,
remove the filter, cap the inlet for moving and return the
instrument to its normal location and operating status.
9.
If the instrument still fails the Purge Test, contact Lighthouse
Tech Support for assistance.
Summary of Contents for REMOTE 3104P
Page 1: ......
Page 2: ...00...
Page 3: ...Lighthouse Worldwide Solutions REMOTE 3104P 5104P Airborne Particle Counter Operating Manual...
Page 6: ......
Page 10: ...Lighthouse REMOTE 3014P 5104P Operating Manual t iv 248083336 1 Rev 3...
Page 12: ...Lighthouse REMOTE 3014P 5104P Operating Manual ii 248083336 1 Rev 3...
Page 40: ...Lighthouse REMOTE 3104P 5104P Operating Manual 4 10 248083336 1 Rev 3...
Page 54: ...Lighthouse REMOTE 3104P 5104P Operating Manual 6 12 248083336 1 Rev 3...
Page 74: ...Lighthouse REMOTE 3104P 5104P Operating Manual B 2 248083336 1 Rev 3...
Page 76: ...Lighthouse REMOTE 3104P 5104P Operating Manual C 2 248083336 1 Rev 3...
Page 81: ...00...
Page 82: ......