Lighthouse REMOTE 3104P, 5104P Operating Manual
A-6
248083336-1 Rev 3
Device Options (40050)
displays the instrument’s device configuration.
If bit 0 of Register 40050 is set, it indicates that the instrument is
capable of Fast Download.
If Bit 1 of Register 40050 is set, the instrument will display and update
the data registers every second. No data will be recorded in the data
buffer. Enabling this feature disables the Print on Sample feature.
If Bit 2 of Register 40050 is set, the value in Register 40052 will be
used to access data in the buffer to be displayed in data registers
3xxxxx. If Bit 2 is 0, the index value in Register 40025 will be used to
access data in the buffer.
If Bit 3 of Register 40050 is set, the number of locations available
(Register 40199) is set to 400 and Registers 40200-40999 are used for
four character names associated with location index values. For
example, the name for location 3 would be located at registers 40204-
40205.
If Bit 3 of Register 40050 is 0, up to 200 locations can be specified.
Registers 40200-40999 are reserved for eight character names
associated with location index values. Thus the name for location=3
would be located at registers 40208-40211. Register 40199 indicates
the number of location names supported on this device.
Registers 40200-40999 are reserved for eight character names
associated with location index values. Thus the name for location =3
would be located at registers 40208-40211. Up to two hundred
locations can be specified.
Register 40199 indicates the number of location names supported on
this device.
Table A-5 Device Options
Bit
Description
0
Fast Download (1=Enabled, 0=Disabled) non-writable
1
One Second Data Update (1=Enabled, 0=Disabled)
2
Display Data Record Using Index/Record (1-Record, 0-
Index)
3
Number of Locations (1-400 locations with four character
names, 0-200 locations with eight character names)
4-15
Reserved
Summary of Contents for REMOTE 3104P
Page 1: ......
Page 2: ...00...
Page 3: ...Lighthouse Worldwide Solutions REMOTE 3104P 5104P Airborne Particle Counter Operating Manual...
Page 6: ......
Page 10: ...Lighthouse REMOTE 3014P 5104P Operating Manual t iv 248083336 1 Rev 3...
Page 12: ...Lighthouse REMOTE 3014P 5104P Operating Manual ii 248083336 1 Rev 3...
Page 40: ...Lighthouse REMOTE 3104P 5104P Operating Manual 4 10 248083336 1 Rev 3...
Page 54: ...Lighthouse REMOTE 3104P 5104P Operating Manual 6 12 248083336 1 Rev 3...
Page 74: ...Lighthouse REMOTE 3104P 5104P Operating Manual B 2 248083336 1 Rev 3...
Page 76: ...Lighthouse REMOTE 3104P 5104P Operating Manual C 2 248083336 1 Rev 3...
Page 81: ...00...
Page 82: ......