Lighthouse REMOTE 3104P, 5104P Operating Manual
4-6
248083336-1 Rev 3
Setting the Real Time Clock
The Real Time Clock (RTC) can be read in registers 40027 and 40028:
In order to change the RTC to the current local date/time, enter the high
and low values as unsigned integers to registers 40035 and 40036
respectively, which are the Data Set registers.
Then write the command
13
to the command register 40002. This will
write the values in the Data Set registers (40035 and 40036) to the RTC
registers (40027 and 40028).
Changing the Default Instrument Parameters
The main instrument parameters involved with the operation of the
REMOTE counter are Location, Sample Time, Hold Time and Initial
Delay.
The Location is set by writing an unsigned integer to register 40026.
The range of values is from 0 to 999.
Sample Time, Hold Time and Initial Delay all use 2 registers, a high
word and a low word. If the desired value for any of these parameters
is less than 9 hours 6 minutes 8 seconds, then only the low word
register needs to be written (in seconds).
Table 4-3 Real Time Clock Registers
Register
Data Type
Description
40027
signed integer
Real Time Clock (RTC) [high]. Works in
conjunction with 40028. Displays date and
time, in number of seconds since
midnight, 1/1/1970.
40028
signed integer
Real Time Clock [low]
Table 4-4 Data Set Registers
Register
Data Type
Description
40035
unsigned integer
Data Set [high]. Works in conjunction
with 40036. Data entered here is applied
to the device through the command
register.
40036
unsigned integer
Data Set [low]
Summary of Contents for REMOTE 3104P
Page 1: ......
Page 2: ...00...
Page 3: ...Lighthouse Worldwide Solutions REMOTE 3104P 5104P Airborne Particle Counter Operating Manual...
Page 6: ......
Page 10: ...Lighthouse REMOTE 3014P 5104P Operating Manual t iv 248083336 1 Rev 3...
Page 12: ...Lighthouse REMOTE 3014P 5104P Operating Manual ii 248083336 1 Rev 3...
Page 40: ...Lighthouse REMOTE 3104P 5104P Operating Manual 4 10 248083336 1 Rev 3...
Page 54: ...Lighthouse REMOTE 3104P 5104P Operating Manual 6 12 248083336 1 Rev 3...
Page 74: ...Lighthouse REMOTE 3104P 5104P Operating Manual B 2 248083336 1 Rev 3...
Page 76: ...Lighthouse REMOTE 3104P 5104P Operating Manual C 2 248083336 1 Rev 3...
Page 81: ...00...
Page 82: ......