Programming
248083336-1 Rev 3
4-7
The low word register for Sample Time is 40034.
The low word register for Hold Time is 40032.
The low word register for Initial Delay is 40030.
Table 4-5 Instrument Parameters
Register
Data Type
Description
40026
unsigned integer
Location number; Specifies location of
Particle Counter.
40029
unsigned integer
Initial Delay [high]. Works in conjunction
with 40030. Number of seconds to wait
before starting the first sample. Max value
is 359,999, which equals 99h 59m 59s.
40030
unsigned integer
Initial Delay [low]
40031
unsigned integer
Hold Time [high]. Works in conjunction
with 40032. Number of seconds to wait
between sample periods. Max value is
359,999, which equals 99h 59m 59s
40032
unsigned integer
Hold Time [low]
40033
unsigned integer
Sample Time [high]. Works in
conjunction with 40034. Number of
seconds to sample. Max value is 86,399,
which equals 23h 59m 59s.
40034
unsigned integer
Sample Time [low]
Summary of Contents for REMOTE 3104P
Page 1: ......
Page 2: ...00...
Page 3: ...Lighthouse Worldwide Solutions REMOTE 3104P 5104P Airborne Particle Counter Operating Manual...
Page 6: ......
Page 10: ...Lighthouse REMOTE 3014P 5104P Operating Manual t iv 248083336 1 Rev 3...
Page 12: ...Lighthouse REMOTE 3014P 5104P Operating Manual ii 248083336 1 Rev 3...
Page 40: ...Lighthouse REMOTE 3104P 5104P Operating Manual 4 10 248083336 1 Rev 3...
Page 54: ...Lighthouse REMOTE 3104P 5104P Operating Manual 6 12 248083336 1 Rev 3...
Page 74: ...Lighthouse REMOTE 3104P 5104P Operating Manual B 2 248083336 1 Rev 3...
Page 76: ...Lighthouse REMOTE 3104P 5104P Operating Manual C 2 248083336 1 Rev 3...
Page 81: ...00...
Page 82: ......