Head
GA10218_0101 - 5/2002
⏐
56
6.4.1 Trigger Level 1
Main menu
>
Trigger & Alarms
>
Trigger Level 1
The value of the first trigger level can be set. See Chapter
4.1.2.7
Numeri-
cal Entries
for the description of the entry.
Trigger 1, 2 and Trigger 3 are programmable switching thresholds. When
these thresholds will be exceeded the PhoeniXL reacts as follows:
Display
In the status line of the display the signs for Trigger 1, 2 and Trigger 3 are
displayed inverted if the leak rate exceeds (becomes higher than) the pro-
grammed value. (see Fig. 14 &15)
Relay Output
The trigger-relais of the digital out switches. Please refer to Chapter
2.2.2.2
, Digital out, for further details.
Alarm/Loudspeaker
Additionally Trigger level 1 defines at which level the various alarm types
react (See Chapter
6.4.7
,
Audio alarm type
)
6.4.2 Trigger Level 2
Main menu
>
Trigger & Alarms
>
Trigger Level 2
The value of the second trigger level can be set. Please refer to Chapter
4.1.2.7
Numerical Entries
for the description of the entry.
If Trigger 2 is exceeded the corrensponding relay will switch. This is also
indicated at the display (see
6.4.1
).
6.4.3 Trigger Level 3
Main menu
>
Trigger & Alarms
>
Trigger Level 3
The value of the second trigger level can be set. Please refer to Chapter
4.1.2.7
Numerical Entries
for the description of the entry.
If Trigger 3 is exceeded the corrensponding relay will switch. This is also
indicated at the display (see
6.4.1
).
6.4.4 Units
Main menu
>
Trigger & Alarms
>
Units
The prefered leak rate unit can be selected. There is the choice of 4 (mbar,
Pa, Torr, atm) pressure units and 5 leak rate units (mbar l/s, Pa m
3
/s,
Torr l/s, atm cc/s, s ft
3
/yr).
In Sniff mode the following measuring units are selectable (Refer to chap-
ter
6.3
): ppm, g/a eq (helium leak rate is equivalent with leak rate R134a),
oz/yr eq (helium leak rate is equivalent with leak rate R134a).