Description of the operation mode of the Digital Out. The pin assignement
for contacts 8 to 16 follows the same order as for pins 5 to 7.
The actual pin setting can be seen under Info/View internal data/page 7.
The following digital out signals are selectable.
Trigger 1:
Is open in case Trigger Level 1 is exceeded or the machine is not in condi-
tion of measuring.
Trigger 2:
Trigger Level 2 analog Trigger 1.
Trigger 3:
Triggerlevel 3 analog Trigger 1
Zero active:
Is closed in case Zero function is running.
Ready:
Is closed in case machine is ready for measurement (Emission on, no
error).
CAL active
Closed when machine is in calibrating routine.
CAL Request
Is opend in case of calibration request. During external calibration a open
output indicates that the external calibrated leak has to be closed.
Fail
Open when a error is shown.
Warning
Open when a warning is shown.
Gas ballast
Closed when gas ballast is active.
Open
Open all time.
Close
Closed all time.
Recorder Strobe
Closed in case recorder output is invalid. Only used when record output is
set on „leak rate“.
6.6.4.5 Define PLC inputs
Main menu > Settings > Interfaces > Define PLC inputs
These inputs can be used to control the PhoeniXL with a programmable
logic control (PLC).
Pin
Assignment
1
PLC in free selectable
2
PLC in free selectable
3
PLC in free selectable
4
PLC GND