9 Maintenance
9.1 LEYBOLD Service
If equipment is returned to Leybold, indicate whether the equipment is
free of substances damaging on health or whether it is contaminated. If it
is contaminated also indicate the nature of the hazard. For this you must
use a Declaration of Contamination form
Fig. 1
which has been prepared
by us which we will provide upon request or you may copy the form which
is attached in the folder of documents. As well you can download this
form from our homepage: www.leybold.com under „Support & Down-
load“.
Please attach this form to the equipment or enclose it with the equipment.
This Declaration of Contamination is required to meet German Law and to
protect our personnel. LEYBOLD must return any equipment without a
Declaration of Contamination to the sender’s address.
Before shipping fir the yellow screw-on seals on to the connections
EXHAUST and GAS BALLAST.
9.1.1 Maintenance Plan
Maintenance work should be done on the PhoeniXL as required. This
work will normally be limited to exchanging the oil in the Trivac D2,5E
rotary vane pump and the built in air and oil filters.
As a preventive measure it is recommended that you check the rotary
vane pump once a month. Here note schould be taken of the oil level and
the colour of the oil.
Attention
Only Arctic oil (Cat. no. 20028181) must be used in the TRIVAC
D2,5E in the PhoeniXL.
The monthly interval for the check is just a nominal period. If the PhoeniXL
is used heavily, in particular in sniffer mode, then this check should be
performed more frequently. The rotary vane pump is located on the side
of the mechanical section at the bottom of the leak detector.