Head
GA10218_0101 - 5/2002
⏐
80
6.6.7.2 Pressure limits for sniff mode
Main menu > Settings > Monitoring functions > Pressure limits sniff
mode
This function is automatically activated in sniff mode. The pressure limits
define an upper and lower limit of the inlet pressure P1. If the pressure is
not in this range error messages are generated:
P > upper limit: Capillary broken
P < lower limit: Flow through capillary too low (Capillary blocked)
Softkey No. 2:
↓
Decrease change over threshold EVAC-GROSS
Chosable between15-3 mbar (Default value 15 mbar)
Softkey No. 3:
↓
Decrease change over threshold GROSS-FINE
Chosable between 0,2-0,05 mbar (Default value0, 2 mbar).
Softkey No. 4
Adjustment for ARGON
Selection between air or Argon
Softkey No. 5:
?
Help text
Softkey No. 6:
↑
Increase change over threshold EVAC-GROSS
Chosable between 3 - 15 mbar
Softkey No. 7:
↑
Increase change over threshold GROSS-FINE
Chosable between 0,05 - 0,2 mbar
Softkey No. 2:
↓
Decreasing the maximum pressure, Upper limit is 0,15 mbar (default)
Softkey No. 3:
↓
Decreasing the minimum pressure
Softkey No. 4:
Setting for Quicktest
Setting for use with the Quicktest
Softkey No. 6
↑
Increasing the maximum pressure
Softkey No. 7:
↑
Increase the minimum pressure
Lower limit is 0,05 mbar (default)