E61 Emission fail.
Emission could not be switched on. MSV subassembly indicates a
fault. MENB emission current not within range.
MSV board is defective.
Both cathodes are defective, replace ion source.
W62 Flow through capillary to low. In the sniffer mode the intake pressure of the sniffer line is controlled.
If the pressure falls below the minimum limit, the flow through the
capillary is too low (contamination) or the capillary is blocked (foreign
objects, particles).
The minimum limit can be set by the menu. Default value is 0.05
mbar.
6.6.1.7
.
Filter in the tip is clogged.
W63 Capillary broken
In the sniffer mode the intake pressure of the sniffer line is controlled.
If the pressure exceeds the maximum limit, the flow through the
capillary is too high (no leak tightness, broken capillary).
The maximum limit can be set by the menu. Default value is 0,15
mbar.
6.6.1.7
.
The capillary is broken or has been torn off.
E64 TMP error: Nomimal speed
has been exceeded by over
10%
Nominal speed of the pump has been exceeded by over 10%
EMC problems: Check connecting cable, insert it properly. Switch
the power supply voltage off and then on again.
Turbo.Drive S faulty: Inform Leybold Service
E65 TMP error: Pass through time
exceeded
Max. time for passing through the critical frequencies has been
exceeded.
Forevacuum or high-vacuum pressure to high: see chapter 6.6.7.1
to reduce the inlet pressure of the PhoeniXL
Bearing defective: Inform Leybold service for repair
E66 TMP error: Bearing tempera-
ture too high (>67°C)
Max. bearing temperature has been exceeded
Forevacuum or high-vacuum pressure to high: see chapter 6.6.7.1
to reduce the inlet pressure of the PhoeniXL
Fan defective: Replace the fan
Ambient temperature too high: Feed cooler air to the PhoeniXL
Bearing defective: Inform Leybold service for repair
E67 TMP error: Short circuit in
TMP-motor or connecting
cable
Short circuit in the pump’s motor or the connecting cable
Check to see if the connecting cable is undamaged, exchange if
required.
Inform Leybold service in case of short circuit in TMP motor
E68 TMP error: temperature con-
verter too high (>75°C)
Maximum temperature for the converter has been exceeded.
Ambient temperature too high: Feed cooler air to the PhoeniXL
Fan defective: Replace the fan
Forevacuum or high-vacuum pressure to high: see chapter 6.6.7.1
to reduce the inlet pressure of the PhoeniXL