![LEYBOLD 250000 Operating Instructions Manual Download Page 13](http://html1.mh-extra.com/html/leybold/250000/250000_operating-instructions-manual_1890883013.webp)
Head
GA10218_0101 - 5/2002
⏐
13
1.3.2.3 Other Data
1.3.2.4 Ambient Conditions
1.3.3 Technical Data PhoeniXL
300 dry
1.3.3.1 Technical Data
Valves
solenoid
Dimensions (L × W × H) in mm
495 x 456 x 314
Weight in kg
40.0
Noise level dB (A)
< 54
max. Audio alarm dB (A)
90
Contamination level (to IEC 60664-1)
2
Overvoltage category (to IEC 60664-1)
II
For use within buildings
Permissible ambient temperature (during operation) +10 °C … +40 °C
Permissible storage temperature
-10 °C … +60 °C
Max.
rel.
humidity
80% (up to 31°C)
linear decreasing to
50% at 40°C
Max. permissible height above sea level
(during operation)
2000 m
Max. inlet pressure
15 mbar
Minimum detectable Helium leak rates
in vacuum mode
≤
3·10
-11
mbar l / s
in sniffer mode
< 1·10
-7
mbar l / s
Minimum detectable Hydrogen leak rates
in vacuum mode
≤
1·10
-8
mbar l / s
in sniffer mode
< 1·10
-7
mbar l / s
Maximum Helium leak rate which can be displayed 0.1 mbar l / s
Measurement range
11 decades
Time constant of the leak rate signal (blanked off,
63% of the final value)
<1 s
Pumping speed (Helium) at the inlet
– GROSS mode
0.02 l / s
– PRECISION mode
0,4 l / s
– FINE mode
> 2.5 l / s
Detectable masses
2, 3 and 4
Mass spectrometer
180° magnetic sector
field
Ion source
2 filaments;
Iridium/Yttria-oxide
Inlet port
DN 25 KF
Run-up time (after starting)
≤
2 min