6.4.6 Alarm delay
Main menu
>
Trigger & Alarms
>
Alarm delay
In some applications (for instance during pump down in a „chamber test
system“) it might be necessary to block an alarm for some time after pres-
sing START.
This delay time of the alarm can be changed.
After pressing START the loudspeaker is activated as soon as the leak
rate drops below trigger level 1 or after the entered alarm delay time has
elapsed. This setting is only active for the audio alarm types SETPOINT
and TRIGGER ALARM (See Chapter
6.4.7
).
6.4.7 Audio alarm type
Main menu
>
Trigger & Alarms
>
Audio alarm type
The audio alarm type can be chosen.
Softkey 3:
↓
Decrease the delay time. The minimum value is 0 seconds.
Softkey 5: ?
Help text.
Softkey 7:
↑
Increase the delay time. The maximum value is 10 minutes up to infinity.
Softkey 2:
Pinpoint
This function is for localization of a known leak rate value. Details see
chapter 6.4.7.1
Softkey 3:
Leak rate prop.
The sound will be proportional to the leak rate signal. Details see chapter
6.4.7.2
Softkey 5: ?
Help text
Softkey 6:
Setpoint
The sound will be proportional to the leak rate signal only if trigger 1 is
exceeded. Details see chapter 6.4.7.3.
Softkey 7:
Trigger alarm
An alarm sounds when the trigger 1 is exceeded. Details see chapter
6.4.7.4.