2 Installation
2.1 Working Location
Caution
Danger of electrical shock. The PhoeniXL must not be operated
while standing in water or when exposed to drip water. The same
applies to all other kinds of liquids.
Attention
Avoid contact with bases, acids or solvents as well as exposure
to extreme climatic conditions.
Attention
The PhoeniXL is designed for indoor use only.
Attention
Ensure a sufficient air cooling. The air inlet as well the air
discharge openings must never be obstructed.
It is recommended that you check all major helium sources in the vicinity
of the PhoeniXL within a radius of about 10 m for the presence of any big
leaks. You may use the sniffer probe for this.
Fig. 2: View of the PhoeniXL
1
inlet flange