- 25 -
Minimum and maximum values which can be set for zero and full scale corrections
:
ANALOG OUTPUT TYPE
Minimum
Maximum
0÷10 V
0
10.200
0÷5 V
0
5.500
0÷20 mA
0
22.000
4÷20 mA
3.5
22.000
NOTE:
the analog output may also be used in the opposite manner, i.e. the weight setting that
corresponds to the analog zero (
) may be greater than the weight set for the analog full scale
(
). The analog output will increase towards full scale as the weight decreases; the analog
output will decrease as the weight increases.
Example:
= 10000
= 0 analog output 0÷10 V
Weight =
0 kg
analog output = 10 V
Weight = 5000 kg
analog output =
5 V
Weight =10000 kg analog output =
0 V
TEST
-
Input test:
: ensure that for each open input
is displayed,
is displayed when the input is closed.
-
Output test:
: setting
ensure that the corresponding output opens. Setting
ensure that the
corresponding output closes.
-
Millivolt test:
: it displays the response signal of each load cell expressed in mV with three decimals.
COMMUNICATION INTERFACE
-
Input test:
close the inputs to be tested and check that INS correctly reports their status.
-
Output test:
force the status of the outputs to be tested in CMDOUT and check that OUTS correctly reports
their status.
-
Millivolt test:
send command 6137 to CMDR, read the value in R1 and divide it by 10000 to get the mV.