- 13 -
MAXIMUM CAPACITY (BASE program)
: maximum displayable weight (from 0 to full scale; default: 0). When the gross weight exceeds
this value by 9 divisions, the display shows
. To disable this function, set 0.
COMMUNICATION INTERFACE
WRITING
READING
Write the value in W1.
Send command 6016 to CMDR.
Send command 6015 to CMDR.
Read the value in R1.
TARE WEIGHT ZERO SETTING
Perform this procedure after having set the THEORETICAL CALIBRATION data.
Use this function to set to zero the weight of the empty system after commissioning and then later
on to compensate zero variations due to the presence of product residues.
Procedure:
-
confirm the message
by pressing
;
-
the weight value to be set to zero is displayed. In this phase all of the LEDs are flashing;
-
confirming once again, the weight is set to zero (the value is stored to the permanent memory);
-
press to display the value of the total weight reset by the instrument, given by the sum of all
of the previous zero settings.
COMMUNICATION INTERFACE
To reset the tare weight, send command 100 to CMDR.
ZERO VALUE MANUAL ENTRY
WARNING:
perform this procedure only if it’s not possible to reset the weighed structure tare, for
example because it contains product that cannot be unloaded.
Set in this parameter the estimated zero value (from 0 to 999999; default: 0).
COMMUNICATION INTERFACE
WRITING
READING
Write the value in W1.
Send command 6044 to CMDR.
Send command 6043 to CMDR.
Read the value in R1.