Proline Kryomats
32
Starting up
YAUE0007 / 2/26/2009
1 s
−
An acoustic signal is emitted for about 1 second.
−
It is quite normal if the refrigerating machine makes a rattling sound
for a few seconds.
−
The unit starts its self-test. All display segments and symbols
appear for about 1 second.
−
The momentary bath temperature is displayed,
−
the pump starts provided "Standby" or "Manual start" (
Ö
not been programmed,
−
all values are accepted which were active before switch-off.
Check or set overtemperature cut-off point:
−
The switching point is shown in the LED display on pressing the key
.
−
Change overtemperature cut-off (
Ö
7.12.1). Overtemperature
protection and checking on page 75.
)
−
If necessary, top up heat transfer liquid, this has been pumped out
by filling the external load.
−
Display for
KDTDK
(low level) appears when the bath has too
little liquid.
−
Red
LED
above the fault triangle
flashes.
−
Find cause of fault and, where necessary, top up missing heat
transfer liquid (
Ö
−
Press the Enter key.
−
Also, press the key if unit has been switched off in the fault state.
−
No release is possible on Command control element!
Command
−
Language
−
If the Command Console is being
switched on for the first time, the
illustrated window appears automatically,
enabling you to select the dialog language
with the appropriate soft key.
Self-test
77(((
°C
EXT
Bath temperature
/1%21
°C
Level alarm
KDTDK
°C
Overtemp. cut-off
8/
°C
English
Deutsch
Français
Español