7.5
Important settings
7.5.1 Temperature setpoint setting
The setpoint is the temperature, which the thermostat should reach and maintain constant.
Master
(main level)
−
RDr
−
Press key until
RDr
(Setpoint) appears.
−
Press key, display flashes.
or
−
Enter the setpoint with the two keys (
Ö
Wait 4 seconds or
−
Display flashes 4 seconds
Æ
new value is automatically accepted,
or
value is accepted immediately with Enter key.
)
−
For safety reasons the setpoint can only be set up to 2°C above
upper limit of the operating temperature range for the relevant
device type.
−
In the following cases, the manual setpoint entry is blocked:
Setpoint is taken from the analog module, from the programmer in
the Command Console or via the serial interface.
−
When the setpoint temperature is to be lowered, it may take up to
one minute before the blue LED
lights.