Proline Kryomats
88
Starting up
YAUE0007 / 2/26/2009
Command Explanation
0 = make-contact open/ 1 = make-contact closed.
IN_DO_02
State of Contact output 2:
0 = make-contact open/ 1 = make-contact closed.
IN_DO_03
State of Contact output 3:
0 = make-contact open/ 1 = make-contact closed.
IN_MODE_00
Keys Master: 0 = free / 1 = inhibited.
IN_MODE_01
Control: 0 = int. / 1 = ext. Pt100 / 2 = ext. Analogue / 3 = ext. Serial.
IN_MODE_02
Standby: 0 = Unit ON / 1 = Unit OFF.
IN_MODE_03
Keys Command: 0 = free / 1 = inhibited.
IN_MODE_04
Setpoint offset source: 0=normal/1=ext.Pt/2=ext.analogue/3=ext.serial.
TYPE Read
equipment
type.
VERSION_R
Read software type of control system.
VERSION_S
Read software type of protection system.
VERSION_B
Read software type of Command.
VERSION_T
Read software type of cooling system.
VERSION_A
Read software type of analogue module.
VERSION_V
Read software type of RS232/485 module.
VERSION_D
Read software type of digital module.
VERSION_M_0
Read software type of solenoid valve (Cooling water).
VERSION_M_1
Read software type of solenoid valve (Automatic refill).
VERSION_M_2
Read software type of solenoid valve (Level controller).
STATUS
Read equipment status 0 = OK, -1 = error.
STAT
Read error diagnosis response:
XXXXXXX
→
X = 0 no error, X = 1 error.
1 Char = error.
2 Char = Alarm.
3 Char = Warning.
4 Char = over temperature.
5 Char = low-level error.
6 Char = high-level error (at adjustment alarm).
7 Char = no external control variable.
RMP_IN_00_XXX
Read a program segment XXX
(response: e. g. 030.00_010.00
Æ
set point temperature 30.00°C, time = 10 min,
tolerance = 5.00°C, pump level = 1).
RMP_IN_01
Read the current segment number.
RMP_IN_02
Read the set number of program runs.
RMP_IN_03
Read the current program run.
RMP_IN_04
Read the program to which further instructions apply.
RMP_IN_05
Read which program is running now (0=none).
LOG_IN_00_XXXX
Query a measuring point XXXX from data logger
(Reply: e. g. 020.00_021.23_030.50 => set point temperature = 20.00°C, bath
temperature = 21.23°C, external temperature = 30.5°C).
LOG_IN_01
Query all measuring points from data logger
As a difference to the command “LOG_IN_00”, a tabulator is used here as
separator instead of ,_’ . The measuring points are separated by CR and LF. The
end is marked by CR LF CR LF.
LOG_IN_02
Query the start time from the data logger
(Reply: e.g. 20_14_12_20
Æ
day 20, 14:12:20).
LOG_IN_03
Query the acquisition interval from the data logger (Reply in seconds).