Proline Kryomats
YAUE0007 / 2/26/2009
Starting up
81
7.12.9 Fault list „Alarms and Warnings“
Alarms
Message Meaning
OsKDT
Pump too fast (low level)
KDTDK
Low level alarm in the level sensor
rDLMO
Overtemperature (t
> tmax)
aK/B
Pump blocked (no rotation)
BE@0K
Command control element connection interrupt
@K 0
Temperature signal of external Pt100 missing
@K 1
Temperature signal of analogue input missing
@K 2
Temperature signal of serial port missing
@K 3
Analogue module: Current input 1 interrupted
@K 4
Analogue module: Current input 2 interrupted
@K 5
Protection system: High bath level
@K 6
Error digital input
@K 7
Refill fail
Warnings in the “Master-Display”
Warnings in the “Safety system”
Message Meaning Message
Meaning
VW 0
Overflow of CAN receipt
VW0/0
Overflow of CAN receipt
VW 1
Watchdog-Reset
VW0/1
Watchdog-Reset
VW 2
til-limitation active
VW0/2
Close to bath overflow
VW 3
tih-limitation active
VW0/3
Bath level is approaching switch off level
or is out of optional range
VW 4
Heatsink temperature is superheated
VW0/4
Heater 1 break
VW0/5
Heater 2 break
VW0/6
Heater 3 break
VW 00
Software version of protection system too old
VW00/
Software version of control system too old
VW 01
Software version of operating system too old
VW001
Software version of operating system too old
VW 02
Software version of heating system too old
VW002
Software version of heating system too old
VW 03
Software version of analogue Interface too old
VW003
Software version of analogue interface too old
VW 04
Software version of RS 232 too old
VW004
Software version of RS 232 too old
VW 05
Software version of contact I/0 module too old
VW005
Software version of contact I/0 module too old
VW 06
Software version of valve 0 too old
VW006
Software version of valve 0 too old
VW 07
Software version of valve 1 too old
VW007
Software version of valve 1 too old
VW 08
Software version of valve 2 too old
VW008
Software version of valve 2 too old
VW 1/
Software version of valve 3 too old
VW01/
Software version of valve 3 too old
VW 10
Software version of pump 0 too old
VW010
Software version of pump 0 too old
VW 11
Software version of pump 1 too old
VW011
Software version of pump 1 too old
VW 12
Software version of pump 2 too old
VW012
Software version of pump 2 too old
VW 13
Software version of pump 3 too old
VW013
Software version of pump 3 too old