Proline Kryomats
56
Starting up
YAUE0007 / 2/26/2009
−
The temperature measurement device
shows the true temperature value (with
glass thermometers consider the
correction where applicable!).
−
Change the display in the adjacent
window to the true value with cursor or
soft keys and accept with
or End ,
−
or quit the window with
without making
changes.
7.6.7 Restoring the works setting of the internal temperature-probe offset
If the offset has been misadjusted unintentionally, the works setting can be restored with this function.
Command
−
Default
−
Open the device parameter menu via
the soft key Menu .
−
With the cursor keys continue to
Æ
Settings
Æ
Calibration
Æ
intern Pt100
Æ
Default
.
−
The window shown adjacent appears.
−
Confirm selection with
.
−
Confirm the control dialog on the right
with
or cancel with
.
−
Return to the measurement window with
End or
.
Help Menu Screen T
set
T
fix
6 7 8 9
0
1 2 3 4
5
Temperature value of the ref.
temp. measurement device:
20.
15
Min: -50.0°C Max: 302.00
Help Menu End T
set
T
fix
intern Pt100
extern Pt100
Calibration
Default
Help Menu End T
set
T
fix
Confirm input!
Enter key: Continue
Escape key: Cancel