Proline Kryomats
54
Starting up
YAUE0007 / 2/26/2009
7.6.4 Restoring works settings
All works settings, apart from the control parameters and the sensor calibration, are restored.
Command
−
Works settings
−
Open the device parameter menu via
the soft key Menu .
−
With the cursor keys continue to
Æ
Settings
Æ
Works settings
.
−
The window shown opposite appears.
−
Master and then
only control par.
int.
is shown as a possible choice.
There are however various possibilities,
which can be selected with
or
:
−
Under
All modules
Master, Command
and all connected modules are reset to
the works setting with
all default
.
−
Under
Master
you have the choice
between:
−
all default
, then all Master settings
are reset,
−
only control para. int.
for the
internal control parameters,
−
only control para. ext.
similar for
external,
−
only miscellaneous
which resets
setpoint, pump level, maximum
current consumption, control to
internal and auto start to "Auto".
−
Under
Command
all command settings
are reset with
All default
.
−
Confirm selection with
.
−
Confirm the control dialog shown on the
left with
or cancel with
.
−
Return to measurement window with
End or
.
Help Menu End T
set
T
fix
All modules
Master
Command
Cool
all default
only control par. int
only control par. ext
only miscellaneous
Help Menu End T
set
T
fix
Confirm input!
Enter key: Continue
Escape key: Cancel