Proline Kryomats
100
Maintenance
YAUE0007
/ 2/26/2009
9.3.4.2.2
Decalcifying the water cooling circuit
At regular intervals of 3 months or longer, the water-cooled condenser must be decalcified or cleaned.
This depends on the hardness of the cooling water and the degree of soiling. Drain according to (
Ö
Required equipment:
−
Two containers of approx. 10 to 20 liters volume.
−
Use a suitable pump (drum pump) or a hose with funnel. Place the funnel as high as
possible so that the device can fill quickly.
−
Fit connecting hoses between container, pump, cooling water inlet and between cooling
water outlet and back to container.
Change containers when
top one is empty.
Acting time:
Continue the pump stage until most of the foamy reaction, usually at the
start, has decayed. Generally, this is achieved after about 15 to 30 minutes.
Decalcifier:
Water with LAUDA Decalcifier LZB 126.
It is essential to follow the safety instructions when handling the chemicals.
Flushing:
Allow at least 30 liters of water to flow through.
9.3.5 Repair information
If you need to send in a unit for repair, it is essential to first contact the LAUDA Service Constant
Temperature Equipment
Ö
)
−
When sending in the unit, ensure that it is carefully and properly packed. LAUDA
cannot be held liable for any damage caused by improper packing.
Standard & Alternative