Proline Kryomats
YAUE0007 / 2/26/2009
Starting up
47
Command
−
Control Variable
−
Open the device parameter menu with the
soft key Menu .
−
With the cursor keys, change further to
Æ
Control
Æ
Control Variable
.
−
intern
is currently active.
−
Select other control variables (only
displayed when present) with
or
and confirm with
or End ,
−
or quit the window with without making
any changes.
or
−
Change temperature probe source:
−
0
for the internal probe,
−
DO
only when an external probe is connected,
−
D@
only when an analog module is connected and configured,
−
DR
only when a serial module is connected and is continuously
receiving actual values from a PC.
Wait 4 seconds
or
−
Display flashes 4 s
Æ
new value is automatically accepted,
or
−
value is immediately accepted with Enter key.
−
If
DO
D@
or
DR
has been selected,
then the green
LED indicates that the control has regulated to
the external temperature signal.
Help Menu
End T
set
T
fix
intern
External Pt100
Analog module
Digital module
Control Variable
01!34
°C