11
Technical data
Table 10
Specification
Unit
Value
Ambient temperature range
°C
5 – 40
Temperature range for storage of the drained device
°C
-20 – 43
Temperature range for transport of the device
°C
-20 – 43
Protection level (IP code) according to IEC 60529
---
IP 21
Degree of pollution according to EN 60664-1 / VDE
0110-1
---
Pollution degree 2
Only non-conductive pollution occurs in the
device. However, temporary conductivity due
to condensation must be expected occasion-
ally.
Protection class for electrical equipment
DIN EN 61140 (VDE 0140-1)
---
1
Class division according to DIN 12876-1
‑
Class designation
---
I
‑
Identification code
---
NFL (suitable for non-flammable liquids)
EMC immunity
---
for industrial electromagnetic environments
EMC emitted interference
---
Class B in accordance with CISPR 11
Working temperature range
The temperature range is determined by the
constant temperature equipment.
‑
with an Integral IN XT
°C
-40 – 90
‑
with an Integral IN P
°C
-40 – 140
Measurable range of the flow rate (depending on
the maximum possible flow rate of the constant tem-
perature equipment)
l/min
0 – 99
Flow control range
‑
minimum
l/min
0.2
‑
maximum
Limitations may arise from the system in the max-
imum flow range (Integral IN, application and tem-
perature).
l/min
70
Flow control accuracy,
determined with Kryo 30 at 20 °C, 20 l/min, 1 bar
l/min
±0.2
Measurement deviation
‑
at 1 l/min
%
±1.3
‑
at 70 l/min
%
±0.2
Thread on the hydraulic connections (exterior thread) mm
M38 x 1.5
Thread on the drain nozzle (exterior thread)
---
G 3/8" A
Filling volume
L
1
Dimensions (width x depth x height)
mm
605 x 420 x 620
V2
Flow controller MID 80
31 / 38