5 Commissioning
5.1
Software versions
In order to operate the flow controller, the constant temperature equipment
must be equipped with the specified version of the software or higher.
If this is not the case, the software for your constant temperature equipment
must be updated.
Table 6
Software
Valid from version
Control system
1.19
Pump system
2.53
Flow module
3.34
5.2 Switch the device on and off
WARNING!
Splashing heat transfer liquid
Eye damage
Always wear suitable safety glasses when working on the
device.
The system, consisting of constant temperature equipment, flow controller,
application and hoses, must be properly filled, vented and degassed.
n
The flow controller is automatically switched on when you switch on the
constant temperature equipment.
n
While starting up the system, the through-flow control and pressure
limitation should be switched
off
and the pump set to
0
as appropriate
until it is completely filled, if permitted by the application. Only then
may the pump be started in filling mode. This can then be adjusted via
the pump level.
n
Read and follow the instructions on filling, venting and degassing in the
operating manual of the constant temperature equipment.
n
If degassing is necessary, carefully degas the entire circuit.
n
Top up the constant temperature equipment with the same heat transfer
liquid, if required.
n
Always check the condition of the system before switching it on.