The pressure indication of the pump on the Integral IN display
switches to the measured pressure level in MID 80.
In the basic window of the constant temperature equipment, the pressure of
the heat transfer liquid at the outlet of the flow controller is displayed in the
direction of the application. Next to it the corresponding measured flow in
the direction of the application.
Please note:
n
Once the flow control has been activated (
Status on
), the pump power
is automatically controlled by the flow controller.
n
The application (e.g. glass apparatus) and hoses must be suitable for the
resulting pressure (maximum pressure of the pump in the Integral XT
7 bar On the Integral thermostat P 10).
If the adjustable maximum pressure is exceeded, the constant tem-
perature equipment stops and reports the
gage pressure
alarm.
n
Install a separate safety valve against gage pressure in the circuit to
protect against operating errors.
Fig. 10: Control menu in the Integral XT and Integral P for the flow controller
Fig. 9: Basic windows Integral XT and Inte-
gral P with flow controller
V2
Flow controller MID 80
19 / 38