Tool:
n
Open-end wrench size 41
n
Open-end wrench size 46
1.
The constant temperature equipment is switched off.
2.
Place two hoses between the flow controller and the constant tem-
perature equipment.
3.
Unscrew the screw caps from the connecting sleeves counterclock-
wise by hand. Keep the screw caps in a safe place for later use.
4.
Connect the inlet of the constant temperature equipment to the
input of the flow controller. The input on the flow controller is labeled
OUTFLOW XT / P.
5.
Connect the outlet of the constant temperature equipment to the
output of the flow controller. The output on the flow controller is
labeled RETURN XT / P.
6.
Screw the union nut onto the connecting sleeve clockwise. Tighten
up the union nut on the connecting sleeve clockwise. While doing so,
brace the connecting sleeve with a second open-end wrench.
1.
The constant temperature equipment is switched off.
2.
Lay the LiBus cable of the flow controller to the constant temperature
equipment in such a way that it cannot be tripped over.
3.
Insert the LiBus cable in the LiBus interface on the constant tem-
perature equipment. Secure the connection by turning the ring on the
connector clockwise.
If a free LiBus interface is not available on the constant
temperature equipment:
Use a T-extender adapter cable (EKS 073).
Connecting the constant temperature
equipment
Fig. 7: Connection side to the constant
temperature equipment
Connecting the LiBus cable