Personnel:
n
Operating personnel
Protective equipment:
n
Safety glasses
The
Drain
submenu is only displayed if a flow controller is connected to the
constant temperature equipment via LiBus.
1.
Place a collecting basin under the drain nozzle on the flow controller
and remove the blind cap.
2.
Screw a hose nozzle with a hose onto the drain nozzle (G 3/8" A).
3.
Run the hose with a slight fall to a sufficiently large collecting vessel.
4.
Press the [Enter key] on the constant temperature equipment to
open the menu.
5.
Select the menu items
Fill mode
Drain
and confirm with the
Enter key.
The options [off] and [on] appear on the display.
6.
Select the option [on] and confirm the entry.
The control valve in the flow controller is moved to the center
position for the optimum drainage of all lines.
7.
Open the drain tap on the flow controller. To do this, turn the tap
counterclockwise. Wait until the flow controller has been completely
drained.
8.
Close the drain tap, unscrew the hose nozzle and screw the blind cap
onto the drain nozzle.
The flow controller has been drained.
9.
Press the [Enter key] on the constant temperature equipment to
open the menu.
10. Select the menu items
Fill mode
Drain
and confirm with the
Enter key.
The options [off] and [on] appear on the display.
11.
Select the option [off] and confirm the entry.
The draining of the flow controller has been completed.
V2
Flow controller MID 80
25 / 38