3 Device description
3.1
Overall view of Integral (large casing version)
1 Filler nozzle with cover
2 Stationary operating unit
3 Interfaces and two slots for interface module
4 Command Touch remote control (optional)
5 Mains switch
6 Adjusting wheel for bypass valve
7 Pump connector
8 Service socket (only for service)
9 Drain tap for expansion vessel, constant temperature equipment
10 Drain tap for low controller
11 Drain tap for the hydraulic circuit, constant temperature equipment
12 Mains cable
13 Connecting sleeve for cooling water (only water-cooled devices)
14 Four castors (front castors with parking brake)
In addition, there is an overflow pipe on the rear of the device (con-
cealed)
3.2 Mains switch
The devices have a mains switch on the right side of the device. This is a
rotary switch. Position [0] switches the device off, while Position [I] switches
it on.
Mains switch for three-phase devices
The automatic fuse may only be switched on by a qualified electri-
cian!
The rotary switch is not designed as a safety switch. The three-
phase devices have a separate automatic fuse built in. If the
automatic fuse trips, contact the LAUDA Constant Temperature
Equipment Service department.
3.3 Hydraulic circuit
The hydraulic circuit in the Integral XT FC consists of a piping system
through which the heat transfer liquid flows under pressure.
All devices are equipped with an 8-stage, hermetically sealed (magnetically
coupled) pump. The pump performance can thus be optimally adapted to the
respective task: High pump pressure, if, for example, long hoses lead to the
external consumer.
A programmable maximum value of the pump pressure allows effective pro-
tection of pressure-sensitive applications by switching off to standby mode
when the preset value is reached.
In addition, either the flow rate can be measured and regulated within a
defined range or the pump pressure can be preset via pressure control.
Fig. 5: View of Integral 1850 XTW FC MID
70
Hydraulic circuit in Integral XT FC
V4
Integral process thermostats with flow control unit
21 / 171