The currently measured flow rate in l/min is displayed in the
Modules
Flow control
menu under the
Through-flow control
menu item. To set the
Chapter 5.13 “Setting the flow rate” on page 80.
1.
Switch off the constant temperature equipment at the mains switch.
This also switches off the flow controller.
2.
Disconnect the constant temperature equipment from the power
supply (mains plug) before carrying out any further installation or
maintenance work.
5.13
Setting the flow rate
The currently measured flow rate is displayed in the
Modules
Through-
flow control
menu under the
Flow rate
menu item. A setting cannot be made
in the
Flow rate
menu item.
1.
Press the [Enter key] on the constant temperature equipment to
open the menu.
2.
Select the menu items
Modules
Through-flow control
Setpoint
.
The entry window opens.
You can see the permissible limit values
Max:
and
Min:
as well as
the set point currently set for the flow rate. The value is provided
in l/min.
3.
Enter the desired set point and confirm the entry with [OK].
The flow controller adjusts the flow to the new set point. It may
take up to one minute before the new set point is reached.
Fig. 45: Flow controller menu
Switching off devices
Fig. 46: Through-flow control menu
Fig. 47: Setting the flow rate set point
V4
Integral process thermostats with flow control unit
80 / 171