Table 37: Interface module and interface commands
Interface
Interface command
ID
Description
Ethernet interface
OUT_MODE_06_1
72
Interface command activates
Safe Mode
Activation is not possible if there is
an error
RS 232/485 interface
OUT_MODE_06_1
72
Interface command activates
Safe Mode
Table 38: Alarms which activate the Safe Mode function
Alarm
Description
Description of action
Alarm 9 External actual value is not
available
Temperature value is not trans-
mitted from the external sensor.
The settings of the temperature setpoint are
accepted from Safe Mode
Configuration
menu.
Control variable is switched automatically to
internal.
Alarm 12 Current interface 1,
interruption
Interruption to the analog module The settings of the temperature setpoint
and control variable are accepted from the
Safe Mode
Configuration
menu.
Alarm 13 Current interface 2,
interruption
Interruption to the analog module
Alarm 15 Fault at the digital input
Fault on the digital input/switching
contact
Alarm 16 Refilling has failed
Heat transfer liquid level is too low.
6.23
Importing and exporting data
6.23.1
Importing data
You can import the following data from a USB stick:
n
Control parameters
n
Analog module config.
n
Contact module config.
If you have previously exported the control parameters, analog module con-
figuration or contact module configuration of a device to a USB stick, you
can use the USB stick to transfer them to another device.
Personnel:
n
Operating personnel
1.
Insert the USB stick in the USB host on the operating unit of the
machine.
2.
Press the Enter key to open the menu.
3.
Select the
Data export/import
Data import
menu items in the
main menu.
The submenu opens.
4.
Select one of the following options:
n
Control parameters
n
Analog module config.
n
Contact module config.
Importing data from a USB stick