Specification
Unit
IN 950 XTW FC
IN 1590 XTW FC
‑
minimum
L
4.8
10.0
‑
maximum
L
17.2
30.6
Pump data 50/60 Hz
‑
maximum discharge pressure
bar
3.1
3.1
‑
Maximum flow rate
l/min
45
45
Through-flow control range
l/min
2 – 45
2 – 45
Noise level 50 Hz
Â
dB(A)
67
63
Noise level 60 Hz
Â
dB(A)
69
65
Weight
kg
218
390
Distance between device and
environment
‑
Front
mm
200
200
‑
Back
mm
200
200
‑
Right
mm
200
200
‑
Left
mm
200
200
Flow control accuracy,
determined with Kryo 30 at 20 °C, 20 l/
min, 1 bar
l/min
±0.2
±0.2
Specification
Unit
IN 1850 XTW FC
Device dimensions width x depth x height
mm
950 x 650 x 1605
Filling volume
‑
minimum
L
8.0
‑
maximum
L
28.6
Pump data 50/60 Hz
‑
maximum discharge pressure
bar
6.0
‑
Maximum flow rate
l/min
65
Through-flow control range
l/min
2 – 65
Noise level 50 Hz
Â
dB(A)
62
Noise level 60 Hz
Â
dB(A)
62
Weight
kg
317
Flow control accuracy,
determined with Kryo 30 at 20 °C, 20 l/
min, 1 bar
l/min
±0.2
‑
Front
mm
200
‑
Back
mm
200
‑
Right
mm
200
V4
Integral process thermostats with flow control unit
151 / 171