4 Before starting up
4.1
Install device
WARNING!
Rolling away or overturning of the device due to incorrect han-
dling
Impact, crushing
Do not tilt the device.
Position the device on an even, non-slip surface with a
sufficient load carrying capacity.
Actuate the castor brake when setting up the device.
Do not place heavy parts on the device.
WARNING!
Danger of overpressure from high ambient temperature
Injury, escape of refrigerant
Note the permitted ambient temperature.
WARNING!
The function of the overflow or drain is disrupted
Electric shock
Overflow hose and drainage hose must each slope down-
ward separately all the way into a collecting vessel.
CAUTION!
Risk of heat transfer liquid escaping
Scalding, cold burns
The temperature and media resistance of the hoses used for
the device overflow must be suitable for the application.
Installation conditions:
n
Irritant vapors may develop during operation of the device, depending
on the heat transfer liquid and working temperature used. Always ensure
that the vapors are adequately extracted.
n
Note the electromagnetic compatibility (EMC) requirements of the
device
Chapter 1.3 “EMC requirements” on page 10.
n
Do not cover the ventilation openings of the device.
n
Further installation conditions apply to the devices. These are specified
in the technical data
Chapter 13.1 “General data” on page 147.
V4
Integral process thermostats with flow control unit
27 / 171