Pressure limitation
The value of the pressure limitation must be lower than the set
maximum pressure.
With the value "0 bar" you deactivate the pressure limitation.
1.
Select the menu item
Pump
Flow control
Pressure limit
Pressure limit
.
The Input window opens.
2.
Use the arrow keys to select a value that lies between 0 bar and the
set maximum pressure.
3.
Press [OK] to confirm the new value.
The pressure limitation is activated. The menu item
Pump
Pressure measurement
Outfl. Press. FC
displays the pressure
currently measured at the output of the flow controller.
5.8 Fill the device with heat transfer liquid.
LAUDA is not liable for damages resulting from the use of unsuitable heat
transfer liquids.
Do not mix different heat transfer liquids together!
Use a funnel if required to fill the device with heat transfer liquid.
DANGER!
Use of unsuitable heat transfer liquid
Fire
Select a heat transfer liquid with a temperature range suit-
able for the application.
WARNING!
Overheating of the heat transfer liquid
Fire
The heat transfer liquid used in the device must be set in
the menu of the device.
Setting the pressure limit
Fig. 37: Pressure limit menu
V4
Integral process thermostats with flow control unit
70 / 171