Model 4500 Premier Dust and Opacity Monitor
76
↵
↵↵
↵↵
Opacity:
0%
Clear Path
Opacity:
0%
100% point
Opacity:
0%
Re-cal complete
↵
↵↵
↵↵
Opacity:
0%
100% point
7.
On the User interface press the ENTER key and wait for the display to read '100% Point'.
8.
At the Transceiver undo the 2 socket head screws and remove the Simulator. Store the
Simulator in the carry case provided until required for Re-calibration.
9.
Offer up the Automatic Zero Point Assembly and tighten the 2 socket head screws.
10. Attach the 5 crimp connectors in the same orientation as previously noted.
11. The instrument is now ready to perform 100% point calibration.
12. Always replace the Automatic Zero Point Assembly
before
pressing the ENTER key for
100% point calibration. The motor that lifts the reflector activates immediately after the
100% point calibration.
IMPORTANT
The instrument will display a fault if the Automatic Zero Point Assembly is not replaced
before continuing with the 100% point calibration.