Model 4500 Premier Dust and Opacity Monitor
125
ERROR
DESCRIPTION
DIAGNOSIS
NUMBER
6
Motor jammed
Zero or span cal motors failed to park.
Check motors and micro-switches.
7
Measure LED fail
Signal LED failed.
Check connections. Replace LED.
8
Flood LED failed
Flood LED failed.
Check connections. Replace LED.
9
ADC over-range
Analogue to Digital Converter value too
high. Check correct aperture fitted on
Retro-Reflector.
Call LAND Instruments International Service Dept.
10
ADC fault
Analogue to Digital Converter failed to convert.
Call LAND Instruments International Service Dept.
11
Gain Drift
Flood LED fault during calibration.
Repeat calibration. If fault persists
call LAND Instruments International Service Dept.
12
Filter blocked
Air Filter pressure switch activated.
Clean/replace air filter.
Check wiring between pressure switch and T-Box.
13
Blower failure
Blower pressure switch activated.
Check blower power supply.
Check blower motor.
Check wiring between pressure switch and T-Box.
14
Negative opacity
Opacity value < -5%
New installation: Check the path length is correct.
Calibration error: Check calibration.
Re-calibrate if necessary.
IMPORTANT
Whenever the Land Service Department is contacted regarding problems with an
instrument, please have available information on the
instrument type
and
serial number
.
This will enable the department to respond more accurately and efficiently.