![Land 4500 MKII Premier Operation And Maintenance Manual Download Page 58](http://html1.mh-extra.com/html/land/4500-mkii-premier/4500-mkii-premier_operation-and-maintenance-manual_3476555058.webp)
Model 4500 Premier Dust and Opacity Monitor
58
Register #
Parameter
Data
Description
Range
Write
Type
Access
110
OP1 Span
Float
CLOP 1 range
0.0 - 3.0
YES
111
Optical Density
optical density
112
OP1 Span
Float
CLOP 1 range
0.0 - 1.0
YES
113
Opacity
opacity
114
OP1 Span
16-bit
CLOP 1 range
0 - 4000
YES
Dust
dust density
115
OP1 Average
Float
CLOP 1 average value
Dependent on
YES
116
units of output
(see registers
49 to 53)
122
OP1 Bits
16-bit
CLOP 1 output bits
scaled
NO
4095 = 20mA
126
OP1 Output
8-bit
CLOP 1 output units*
YES
133
OP2 Span
Float
CLOP 2 range
0.0 - 3.0
YES
Optical Density
optical density
134
YES
135
OP2 Span
Float
CLOP 2 range
0.0 - 1.0
YES
Opacity
opacity
136
YES
137
OP2 Span Dust
16-bit
CLOP 2 range
0 - 4000
YES
dust density
138
OP2 Average
Float
CLOP 2 average value
Dependent on
YES
139
units of output
(see registers
49 to 53)
145
OP1 Bits
16-bit
CLOP 2 outputs bits
scaled
NO
4095 = 20mA
149
OP2 Output
8-bit
CLOP 2 output units*
YES
166
Start Calibration
16-bit
Start Calibration
Normally 0
YES
1 - Initiate Cal.
167
Output Relays
16-bit
Output relays*
-
NO
168
Calibration Phases
16 bit
Calibration Phases
-
NO
169
Password
For registers marked * see over page
Note
If the calibration is correctly initiated, the ‘Start Calibration’ register will automatically reset to 0. If the
instrument is unable to initiate the calibration (the instrument is in maintenance mode or already in
calibration) the register will remain at 1 and the calibration will remain pending. Setting the ‘Start
Calibration’ register to ‘0’ will remove any pending operation.