![Land 4500 MKII Premier Operation And Maintenance Manual Download Page 123](http://html1.mh-extra.com/html/land/4500-mkii-premier/4500-mkii-premier_operation-and-maintenance-manual_3476555123.webp)
Model 4500 Premier Dust and Opacity Monitor
123
Termination Box
Modbus Interface:
RS232 or RS485 Multi-drop
Opacity, Dust, Optical Density, Status, Averaging
and Range Information available
Calibration sequence trigger
Analogue Outputs :
Two 4-20mA current loops, fully isolated 50 V d.c.
Independently configurable as: Opacity, Dust,
Optical Density or Calibration Drift
Relay Outputs:
System OK, Span Calibration, Zero Calibration,
Alarm 1, Alarm 2, Maintenance
Relay Rating:
Single Pole Changeover
2A at 40 V d.c.; 2A at 230 V a.c.
Standard Termination Box
Power Supply:
90 to 260 V a.c. (universal input), 50 to 60 Hz
Power Consumption:
10 W
Heated Termination Box
Power Supply:
115 / 230 V a.c. (dual voltage, selectable) 50 to 60 Hz
Power Consumption:
60 W
Installation Category:
Category II
Electromagnetic Compliance:
Conforms to EN50 081-2; EN50 082-2
Operating Temperature:
-20 to +55°C (-4 to +131°F)
Dimensions (H x W x D):
180 x 400 x 120 mm (7 x 16 x 5 inches)
Weight:
7.5 kg (16.5 lb)